15 Wildcat/Wildcat 450 Owners Manual
Troubleshooting
The diagnostic flash codes for your power wheelchair are designed to help you perform basic
troubleshooting quickly and easily. A diagnostic flash code flashes from all 6 of the LED’s
located on the controller in the event one of the conditions listed below develops.
Table 1: Power Wheelchair Controller Internal Diagnostics
Flash
Code
Description Recommended Action
1
Motor Current Over Limit
Motor current is over the controller set limit.
Release the joystick to neutral and try again
2
Battery
Try charging the batteries.
Batteries need replacing.
Check the battery cables and connections
3
Left Motor Check left motor cables and connections
4
Right Motor Check right motor cables and connections
5
Left Parking Brake Check left parking brake cables and connections
6
Right Parking Brake
Check right parking brake cables and
connections
7
User Interface Display PCB fault. Consult a service agent.
8
Controller Consult a service agent.
9
A-series Bus
Communication
Consult a service agent.
If more than one fault exists, the fault having the highest priority (and lowest number of
flashes) is indicated. All faults disable the controller and require the controller to be turned off
then on again once the source of the fault is removed.